►ZOLL AED Pro Surepower Rechargeable Lithium Ion Battery Pack 8019-0535-01 is Suitable for ZOLL AED Pro, ZOLL R and Zoll E Series defibrillators.
ZOLL-Sure-Power-Battery-Pack:
Battery care:
Keep a fully charged spare battery pack with the device at all
times. When the unit displays the message CHANGE BATTERY,
immediately replace the battery pack with a fully charged one.
Regular use of a partially charged battery pack without fully
recharging it between uses might permanently reduce the
battery’s capacity and result in early failure. Regularly test
rechargeable battery packs. A rechargeable battery pack that
does not pass its test could fail without warning. Do not
disassemble a battery pack or dispose of it in fire. Do not try
to recharge a non-rechargeable battery pack. If mistreated, a
battery pack might explode. Dispose of battery packs in
accordance with federal, state, and local regulations. Battery
packs should be shipped to a reclamation facility for recovery
of metal and plastic compounds as the proper method of waste
management. The SurePower Battery System, complete with SurePower Manager Software, Charging Station and SurePower Battery packs–with longer cycles and runtime indicators–provides information to help understand your defibrillator battery demands. This system offers unmatched value in battery management.
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
►zoll-aed-surepower-battery-charger-station-8050-0030-01 $2,275.00 ZOLL AED Battery Power Charging System: Rapid Recharge :Puts batteries back in service fast and interchangeability provides virtually unlimited portable capability with spares.
The ZOLL SurePower Charger
Station is a four-bay unit that can test,
recalibrate, and charge up to four ZOLL rechargable
defibrillator batteries at once. It is a multi-chemistry charger
that works with both ZOLL lithium ion and lead acid batteries. ZOLL SurePower Charger for external battery charging and
capacity evaluation. You can charge up to four battery
packs simultaneously; testing is automatic. ZOLL’s defibrillator batteries and the bays in the SurePower Charger Station are keyed to
prevent incompatible batteries from seating in the charging bays and damaging the SurePower
Charger Station. Place the SurePower Charger Station on a stable, secure surface.
Position the charger so that the control panel and LED’s face the area from which you’ll be working -- this will make it easier to see the LED’s and install and remove batteries.
Make sure that there are at least four inches of clear space surrounding the charger. This
provides the air circulation necessary for the charger to dissipate the heat that it generates
during operation.
A Sure Power Battery Pack Guide Zoll-Sure-Power-Charger-Station-Charging-Bays
http://www.zoll.com/Search.aspx?searchtext=batteries
http://www.zoll.com/uploadedFiles/Public_Site/formtest/Batteries_Sold_Separately.pdf
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Zoll AED Pro Non Rechargeable Lithium Battery Pack 8000086001
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Managing Rechargeable Battery Packs Rechargeable battery packs require a full
recharge after each use. Avoid using a battery pack that is not fully recharged.
results in permanently reduced capacity and early failure of the battery pack. Many factors contribute to the loss of battery capacity, including the frequency of use, the number of battery packs available for operations, and the pattern of discharging and recharging battery packs. Because of this, ZOLL recommends that operators schedule a preventive routine for replacing and discarding used battery packs. Base your replacement schedule for rechargeable battery packs on anticipated use patterns, battery pack testing results, and experience with the device in actual operation. When stored and not in use, a battery pack can lose 2% to 3% of its energy per month. ZOLL recommends purchasing new batteries every eighteen months or sooner. Recharging and Testing Battery Packs ZOLL battery packs are designed to be charged in ZOLL chargers. ZOLL recommends that you always have a ZOLL auxiliary battery charger available for charging spare battery packs and for routine testing of battery packs. WARNING! Test battery packs regularly. A battery that does not pass its test might shut down unexpectedly. For information on using battery chargers, refer to the ZOLL Base PowerCharger manuals listed in “Related Manuals” on page vi. Achieving Optimal Performance with Rechargeable Batteries To ensure the longest life from rechargeable battery packs, follow these general practices: DO carry a fully charged spare battery pack at all times. DO charge battery packs completely. Whenever a battery pack exchange is required, install a fully charged battery pack. Use of partially charged battery packs might result in very short run time and unexpected shutdown. If you use a partially charged battery pack, fully charge that battery pack before using it again. Repeated use after partial charging quickly diminishes the battery pack’s capacity, thereby shortening its life. If you must frequently use partially charged battery packs, your organization should assess whether enough battery packs are available to fully support typical rescue activity. DO implement a means of indicating the charge status of battery packs. It is important to visually distinguish battery packs that are charged from those that are not. Establish a system for visually indicating whether a battery pack requires charging or is charged and ready for use. ZOLL can provide battery pack status labels for this purpose, or you can use labels or methods of your own. DO change the battery pack as soon as the message CHANGE BATTERY appears. The battery warning ultimately leads to shutdown of the unit. As a battery ages, its remaining operating time between the warning and shutdown progressively diminishes. Older battery packs might provide very little run time after the warning and might fail unexpectedly. Always replace a battery with a fully charged battery as soon as the battery warning appears. DO test rechargeable battery packs regularly. Your organization must determine and implement an appropriate testing schedule for rechargeable battery packs. Conformance to this schedule is crucial for identifying battery packs that have reached end-of-life and should be removed from use. Battery packs subjected to repeated short discharge-and-charge cycles might lose capacity quickly, and so should be tested more frequently. DO exchange battery packs regularly. Exchange battery packs once every shift or once every day, depending on their use. DO exercise or test battery packs every 90 days (or sooner as they age). APPENDIX B RECHARGEABLE BATTERY PACKS DO NOT store battery packs in a discharged or depleted state. When you remove a rechargeable battery from the unit, immediately place it in a charger or test well. Idle battery packs lose some of their charge and might suffer damage to charge capacity if left in a discharged state. DO NOT assume that a shift check of the unit verifies adequate battery pack run time. Test the unit daily to verify its readiness for use. This test, however, does not verify adequate charge state or capacity of the battery pack and might leave the unit with inadequate run time. If the message CHANGE BATTERY appears during testing, replace the battery pack immediately. If the removed battery pack is rechargeable, recharge it immediately. DO NOT charge battery packs at temperature extremes. Charge battery packs at or near normal room temperature (15°C to 35°C or 59°F to 95°F). DO NOT remove a partially charged battery pack from the battery charger. Always fully charge a battery pack before returning it to use. If you must use a partially charged battery pack, be sure to fully charge the pack before its next use.
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
►zoll-aed-pro-smart-rechargeable-battery-pack-8004-0103-01 ....$210.00 The ZOLL XL Battery is a 4.5 AH lead acid battery pack intended for use in appropriately configured M Series and M Series CCT defibrillators. It is available as a Smart Battery or Smart Ready Battery. Please see the Smart Battery Operator’s Guide for additional information on Smart NOTE: The XL battery is NOT intended for use in ZOLL 900/1200/1400/1600/1700/2000 series units. See the "Defibrillator Compatibility" section below for information on M Series use. • Follow all recommended maintenance and operating instructions for the XL Battery and associated ZOLL charging/testing systems. If a problem or question arises, contact Technical Service. Do not and you have been adequately trained in the proper use of the XL Battery. • Locate the XL Battery and ZOLL resuscitation devices where they cannot harm the patient or user • Under certain conditions, it is possible for small amounts of hydrogen gas to be released during charging of the battery pack. This may present a hazard. • Do not immerse or set the XL Battery in fluid. • Do not use alcohol or ketones (MEK, acetone, etc.) on the XL Battery. Do not autoclave the XL • The XL Battery is protected against interference from radio frequency emissions typical of two way radios and cellular phones (digital and analog) used in emergency service/public safety activities. Users of the XL Battery should assess the device's performance in their typical environment of use for the possibility of radio frequency interference from high power sources. • Keep a fully charged spare battery pack available at all times with the Monitor/Defibrillator. Replace the battery immediately upon low battery message. • Do not install batteries into Monitor/Defibrillators when storage may exceed 90 days. Battery • Do not use the XL Battery if the case is cracked or broken. • When charging a XL Battery pack in a Base PowerCharger4x4 or Base PowerCharger1x1 unit, ensure unit is labeled “XL Battery Ready” otherwise the battery cannot be charged to full capacity and operating life may be reduced. • ZOLL charging/testing systems used with the XL Battery should be calibrated or tested periodically using the appropriate Service Manual Test Check Out procedures. to a reclaimation facility for the recovery of metal and plastic compounds as the proper method of • Use of ZOLL PD 4410 batteries with M Series CCT or M Series units configured to operate with XL batteries is not recommended. Although the PD 4410 battery can be used to power such units in an emergency situation, the period from low battery warning to unit shut down will be substantially shortened. See the M Series Configuration Guide (9650-0201-01) for additional information. General Information Defibrillator Compatibility M Series units must have software version 30.00 or higher and be outfitted with the appropriate handle assembly. New M Series units may be ordered with the XL Battery option. Older M Series units may be upgraded if the current software version is 20.00 or higher. All M Series CCT units are XL Battery-ready. The XL Battery is not intended for use in ZOLL 900/1200/1400/1600/1700/2000 series units. Charger Compatibility The XL Battery is only compatible with ZOLL Base PowerCharger4x4 or Base PowerCharger1x1 units labeled “XL Battery Ready.” Older ZOLL Base PowerCharger4x4 or Base PowerCharger1x1 units may be upgraded. All XL Battery ready chargers are backward compatible with the PD4410 battery pack. Battery Maintenance Test batteries regularly. At a minimum, ZOLL recommends testing batteries every three months. For additional information on battery maintenance, refer to the ZOLL document, Battery Management Program (#9650-0019). Environmental Specifications Operating Note The XL Battery may feel warm to the touch following an excessive number of defibrillation discharges at maximum energy. Charge Time The XL battery charge times are as follows: Cleaning Clean the ZOLL XL Battery with a soft damp cloth and mild soap. CAUTION! Do not immerse or set battery in fluids; do not use alcohol, ketones (MEK, acetone, etc.); and do not autoclave Upgrade Kits M Series units with software version 20.00 and higher may be upgraded for use with the XL Battery. If the unit is equipped with an Extreme Pack II, an upgrade will be required for the Extreme Pack as well. All M Series CCT units are compatible with the XL Battery. Base PowerCharger 4x4 and Base PowerCharger1x1 units not labeled “XL Battery Ready” must beupgraded for use with the XL Battery.
zoll-aed-pro-smart-rechargeable-battery-pack-8004-0104-01.html
$160.00
.
ZOLL AED PLUS Information Center
|
||||||||||||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||||||||||
|
Saving
Lives Is Our Only Business. We Help Save Lives.
To help our customers
with the very best prices, we buy in large lots from our AED
manufacturers. Therefore, we are able to pass on the
additional, super savings to you. We make owning an AED
as easy as possible because it's a life-saving device. Our
specialist team is the best in the Automated External
Defibrillator industry. Please feel free to call them
any time Toll Free: 1-888-541-2337 for your
Defibrillator needs . Always
Lowest Prices Guaranteed?
why
Defibrillator Manufacturers recommend, you
Should Always
Buy From An Authorized
Automated External Defibrillator
Dealer Only?
why |
||||||||||||||||||||||||||||||||||||||||||||||
| Better Business Bureau Accredited Business. Member Palatine, Illinois (u.s.a) , Chamber of Commerce | ||||||||||||||||||||||||||||||||||||||||||||||
|
Corporate Offices:
348 W. Colfax St., Palatine IL 60067 (U.S.A.) |